User manual

lpErrorDoc.txt
While executing the internal I/Q calibration, a failure occurred.
An internal software error has occurred. A measurement node
requested by the firmware does not exist. A firmware patch is
required.
517 Calibration failure; IQ Calibration failure. IQ Gain iteration error.
{name}
While executing the internal I/Q calibration, a failure occurred.
The IQ Gain could not be calibrated in a reasonable number of
iterations. Please check the IQ Mux hardware.
517 Calibration failure; IQ Calibration failure. Maximum DAC value
exceeded. {name}
While executing the internal I/Q calibration, a failure occurred.
The minimization algorithm cannot provide an accurate calibration.
Correction of the measured error requires a DAC correction greater
than the maximum value of the DAC. Please check the IQ Mux, IQ
Modulator and output board hardware.
517 Calibration failure; IQ Calibration failure. Measurement not stable.
{name}
While executing the internal I/Q calibration, a failure occurred.
The minimization algorithm cannot provide an accurate calibration.
The measurement is not stable which has resulted in an excessive
number of measurement iterations. Please check the IQ Mux, IQ
Modulator and output board hardware.
517 Calibration failure; IQ Calibration failure. Minimum DAC value
exceeded. {name}
While executing the internal I/Q calibration, a failure occurred.
The minimization algorithm cannot provide an accurate calibration.
Correction of the measured error requires a DAC correction less
than the minimum value of the DAC. Please check the IQ Mux, IQ
Modulator and output board hardware.
517 Calibration failure; IQ Calibration failure. Resolution error. {name}
While executing the internal I/Q calibration, a failure occurred.
The minimization algorithm does not have the necessary resolution
to provide an accurate calibration value. Please check the IQ Mux,
IQ Modulator and output board hardware.
517 Calibration failure; IQ Calibration failure. The I channel reference
voltage is out of range. {name}
While executing the internal I/Q calibration, a failure occurred.
The measured Path 2 I channel reference voltage is outside expected
limits. Please check the IQ Mux hardware.
517 Calibration failure; IQ Calibration failure. The Q channel reference
voltage is out of range. {name}
While executing the internal I/Q calibration, a failure occurred.
The measured Path 2 Q channel reference voltage is outside expected
limits. Please check the IQ Mux hardware.
517 Calibration failure; IQ Calibration failure. {name}
Page 52