While executing the internal I/Q calibration, a failure occurred
while attempting to access the IQ CAL CONSTS Array. This array is
required for proper operation of the internal IQ Calibration. This
may be caused by faulty motherboard flash but can also be the
result of a configuration failure during a firmware upgrade or
motherboard replacement.
517 Calibration failure; IQ Calibration failure. Cannot Access IQ CAL
TOLERANCE Array. {name}
While executing the internal I/Q calibration, a failure occurred
while attempting to access the IQ CAL TOLERANCE Array. This array
is required for proper operation of the internal IQ Calibration.
This may be caused by faulty motherboard flash but can also be the
result of a configuration failure during a firmware upgrade or
motherboard replacement.
517 Calibration failure; IQ Calibration failure. Cannot Access IQ INDEX
Array. {name}
While executing the internal I/Q calibration, a failure occurred
while attempting to access the IQ INDEX Array. This array is
required for proper operation of the internal IQ Calibration. This
may be caused by faulty motherboard flash but can also be the
result of a configuration failure during a firmware upgrade or
motherboard replacement.
517 Calibration failure; IQ Calibration failure. Cannot Access IQ20 INDEX
Array. {name}
While executing the internal I/Q calibration, a failure occurred
while attempting to access the IQ20 INDEX Array. This array is
required for proper operation of the internal IQ Calibration. This
may be caused by faulty motherboard flash but can also be the
result of a configuration failure during a firmware upgrade or
motherboard replacement.
517 Calibration failure; IQ Calibration failure. Excessive iterations
during calibration. {name}
While executing the internal I/Q calibration, a failure occurred.
The calibration was unable to optimize Gain, Offset and Quadrature
within a reasonable number of iterations. Please check the IQ Mux,
IQ Modulator and/or output board hardware.
517 Calibration failure; IQ Calibration failure. Failure during RF
calibration. {name}
While executing the internal I/Q calibration, a calibration error
occurred. The RF calibration of I/Q Gain and/or Offset calibration
failed. Please check the IQ Mux, IQ Modulator and output board
hardware.
517 Calibration failure; IQ Calibration failure. Insufficient range. {name}
While executing the internal I/Q calibration, a failure occurred.
The minimization algorithm cannot provide an accurate calibration.
The adjustment DAC does not have the necessary range to remove the
measured error. Please check the IQ Mux, IQ Modulator and output
board hardware.
517 Calibration failure; IQ Calibration failure. Internal software defect.
{name}