In 8340/8757 compatibility mode, the OA query was received but it
is not supported for the current active function.
-100 Command error; R0: No standby mode allowed.
In 8656/57 compatibility mode, R0 was received via GPIB. This
command is not supported by the compatibility mode.
-100 Command error; Remote active function DN/UP not available.
In 8656/57 compatibility mode, either DN or UP was received via
GPIB. These commands are not supported by the compatibility mode.
-100 Command error; Remote active function IS not available.
The instrument only supports the Increment Set (IS) command for
Frequency (FR) and Amplitude (AP) in compatibility mode. All other
uses of IS are invalid.
-100 Command error; Unable to set new amplitude with UP/DN command.
The instrument is unable to obtain the necessary information to
calculate the new amplitude in compatibility mode with the UP/DN
command.
-100 Command error; Unable to set new frequency with UP/DN command.
The instrument is unable to obtain the necessary information to
calculate the new frequency in compatibility mode with the UP/DN
command.
208 I/O error; Could not access BBG{value}'s PRAM file {name}. The
personality has been turned off.
Deleting or copying AUTOGEN_PRAM files while they are in use can
lead to this error. Turning the personality back on should fix
this. If not, report this error to the nearest Agilent Technologies
sales and service office.
208 I/O error; Unable to read bit file.
There is an error reading the specified bit file. Try selecting the
file again.
208 I/O error; The user file is too large.
The user file selected is too large. Try selecting a smaller file.
DCH files must be less than or equal to 4096 bytes. TPC pattern
files must be less than or equal to 256 bytes.
208 I/O error; The user file is too small.
The user file selected is too small. Try selecting a larger file.
208 I/O error; {name}
Invalid file or device name.
209 Invalid argument; {name}
A file system operation failed because an invalid parameter was
specified for the operation. Report this error to Agilent
Technologies.