is defaulted back to EDGE modulation.
-221 Settings conflict; When Skew Mode is enabled, the maximum number of
filter symbols allowed is 32. The filter type has been defaulted.
*** No documentation ***
-221 Settings conflict; When the clock rate setting is below 10 MHz or
above 200 MHz, only phase settings of 0 and 180 degrees are valid. The
phase was changed to a valid setting.
When the frequency is changed to below 10 MHz or above 200 MHz and
the phase is set at 90 or 270 degrees, the phase value is reduced
by 90 degrees. Only 0 and 180 degrees are valid in the frequency
ranges below 10 MHz and above 200 MHz.
-221 Settings conflict; You may not set Forward Link values while in
Reverse Link. Reset Link Direction to proceed.
Values or settings for forward link were entered while in reverse
link. If forward link is the desired link direction, set the ESG to
forward link first, and then re-enter the values or settings.
However, if reverse link is the correct link direction, consult the
SCPI or key references for the correct values or settings to enter.
-221 Settings conflict; You may not set Reverse Link values while in
Forward Link. Reset Link Direction to proceed.
Values or settings for reverse link were entered while in forward
link. If reverse link is the desired link direction, set the ESG to
reverse link first, and then re-enter the values or settings.
However, if forward link is the correct link direction, consult the
SCPI or key references for the correct values or settings to enter.
-221 Settings conflict; You may only set up channels that are part of your
current radio configuration.
An attempt was made to enter a value or setting for a channel that
is not supported in the current radio configuration. The R-FCH and
R-SCH are supported in all the radio configurations for reverse
link. However, the values or settings allowed for these channels
vary according to the radio configuration used. Refer to the Option
201 User's and Programming Guide to determine the supported
channels by radio configuration.
-221 Settings conflict; The {name} packet #{value} subpacket #{value}
should not be overlapped by another subpacket.
Each subpacket should not be overlapped in the channel. Check the
slot position again.
-220 Parameter error
A program data element related error has occurred. This
particular error message is used if the device cannot detect
a more specific error described for errors -221 through -229.
-220 Parameter error; Channel Code exceeds spread factor.
The specified channelization code is too large. The channelization
code should always be at least one less than the spread factor.
-220 Parameter error; Data rate is not valid