match the FM1/PM1 deviation.
-221 Settings conflict; FM2/PM2 value set greater than FM1/PM1 value.
FM1/PM1 changed to match FM2/PM2 value.
FM2/PM2 deviation must be FM1/PM1 deviation. This error is reported
when both FM1/PM1 and FM2/PM2 are enabled and an adjustment to
either FM2/PM2 deviation causes the FM2 or PM2 deviation to be
greater than the FM1 or PM1 deviation. It is also be reported as
FM2/PM2 is turned on, and the last FM1/PM1 deviation setting is
less than the current FM2/PM2 deviation setting. In both cases the
FM1/PM1 deviation is adjusted to match the FM2/PM2 deviation.
-221 Settings conflict; FM/PM deviation decreased to account for FM rate
and ramp sweep settings.
The deviation and rate settings for FM and PM are interrelated when
ramp sweep is active. The deviation setting will be automatically
set to an acceptable value when the rate moves outside an
acceptable range.
-221 Settings conflict; FM/PM rate increased to account for deviation and
ramp sweep settings.
The deviation and rate settings for FM and PM are interrelated when
ramp sweep is active. The rate setting will be automatically set to
an acceptable value when the deviation moves outside an acceptable
range.
-221 Settings conflict; The format must be on and fully configured for use
before being specified as the secondary frame. The secondary frame is
unchanged.
The secondary frame setup must be completely configured as a
stand-alone setup before being selected for use as a secondary
frame. It must have a format, framing, and at least one timeslot
turned on.
-221 Settings conflict; Frame repeat changed to continuous.
Single is not a valid repeat setting for an external data source.
-221 Settings conflict; Frequency list and power list are of unequal size.
Turn one off, set one to size one, or make their sizes equal.
The frequency list has more than one element and the power list has
more than one element, and they are not of equal size. If any of
the frequency, power, or dwell lists have more than one element,
they must all have the same number of elements. A list of a single
element is the same as a list of equal size with the single element
repeated the necessary number of times.
-221 Settings conflict; Frequency Ramp Sweep and the downconverted path
cannot be on at the same time. {name}
The downconverter path below 3.2 GHz feature and Frequency Ramp
Sweep cannot be used at the same time. This error is reported when
the user attempts to activate one feature while the other one is
activated. If Frequency Ramp Sweep is on and the downconverted path
is enabled, Ramp Sweep will be turned off and Step Sweep will be
activated. If the downconverted path is on and Ramp Sweep is
enabled, the lowband path will be set to normal.
Page 27