lpErrorDoc.txt
/ NumberOfBlks. But block size does not match the BlockSetSize
exactly. Try another value.
Block size has been adjusted to match
BlockSetSize / NumberOfBlks. But block size is not match to the
value exactly, please try another value.
-221 Settings conflict; Block size has been adjusted to match BlockSetSize
/ NumberOfBlocks.
Block size has been adjusted to match BlockSetSize /
NumberOfBlocks.
-221 Settings conflict; Block size of DCH{value} must match to
BlockSetSize / NumberOfBlks.
Block size must match to BlockSetSize / NumberOfBlks. The pointed
out DCH should be re-checked.
-221 Settings conflict; Burst cannot be on while using AM.
There is a hardware conflict for the burst envelope (I/Q menu) and
AM1/AM2. The most recently requested modulation will be turned on.
The previous modulation will be turned off.
-221 Settings conflict; Can't find E-DPDCH physical code set.
3gpp standard does not allow the E-DPDCH configuration. Reduce
E-TFC Index, reduce E-DPDCH maximum channelization codes, or
inclease E-DPDCH PL non-max value. See 3GPP TS 25.212 4.8.4.1
Determination of SF and number of PhCHs needed.
-221 Settings conflict; Can't turn on DPDCH.
DPDCH state should be off when Nmax-dpdch is 0.
-221 Settings conflict; Can't turn on HS-DPCCH.
HS-DPCCH state should be off when HS-DSCH Configured value is 0.
-221 Settings conflict; Cannot have more than 4 EDGE multiframe timeslots
on simultaneously. The selected timeslot has been turned off.
An unsupported timeslot type was encountered. EDGE doesn't support
more than 4 multiframe timeslots being turned on.
-221 Settings conflict; Cannot have uplink protocols while the control
frame is on. Frames 1-17 timeslots changed to downlink.
Indicates that, in TETRA mode, some of the timeslots are set to
uplink. When the control frame is turned on in TETRA, all timeslots
in all frames must use downlink protocols. The instrument enforces
this by reconfiguring the timeslots that are not using downlink
protocols when the control frame is switched on.
-221 Settings conflict; Cannot select uplink protocols while the control
frame is on. Selection changed to custom.
Indicates that in TETRA mode, an attempt has been made to select
uplink protocols while the control frame is switched on. TETRA
requires that the control frame be switched off in order to select
uplink protocols.
Page 21