The sweep/list manual point has been reassigned to a smaller number
value due to the longest list decreasing in size or being turned
off. Its new value is the length of the longest enabled list
(frequency or power).
-222 Data out of range; Manual point exceeds power list size. Limiting to
maximum point.
The sweep/list manual point has been reassigned to a smaller number
value due to the longest list decreasing in size or being turned
off. Its new value is the length of the power list, that is the
longest enabled list.
-222 Data out of range; Number of parameters is not an even number
The user power correction calibration points must be an even
number.
-222 Data out of range; OCNS channel(s) power clipped to minimum limit.
As DPCH channel one power is increased, one or more of the OCNS
channels may become clipped to its minimum power of -40dBm during
channel balancing.
-222 Data out of range; Only DEFAULT or USER FILE data type is allowed.
Data type is changed to DEFAULT
Only Default data or user file is valid for Paging channel. Paging
data is now set to Default.
-222 Data out of range; Sequence point exceeds size of active lists.
Clipped to maximum list element.
An error that occurs when in list sweep mode. The index used for
determining the next point to sweep is beyond the valid range for
the current list of points.
-222 Data out of range; Synthesizer: Frequency out of bounds.
The signal generator received an internal request for a frequency
outside of its supported frequency range. Report this error to
Agilent Technologies.
-222 Data out of range; Value clipped to available memory.
There was insufficient memory, so a value was clipped to allow
waveform generation with the available memory. For example, OSR.
-222 Data out of range; value clipped to lower limit.
An input value is below the minimum value allowed. Examples are:
frequency setting, reference, or offset; output power; power
reference and offset; modulation depth, deviation, or modulation
source frequency; number of points and start/stop values for list
mode; start/stop values for internal I/Q calibration; sequence or
register values (save/recall); dwell time.
-222 Data out of range; value clipped to upper limit.
An input value is above the maximum value allowed. Examples:
frequency setting, reference, or offset; output power; power
reference and offset; modulation depth, deviation, or modulation