User manual

lpErrorDoc.txt
-259 LAN setup error; IP address and default gateway must be on the same
subnet. Offending entries are changed.
The IP address and default gateway are not on the same network as
specified by the subnet mask. Verify that both addresses are on the
same subnet and try again.
-259 LAN setup error; IP address is inconsistent with subnet mask or IP
class rules. Offending entry is changed.
The IP address is invalid. Check the entry for syntax and numerical
errors and try again. Verify that the subnet mask is consistent
with this address (host bits should not be all 0s or 1s). If no
subnet is specified, verify that the address is valid with respect
to the default class netmasks: A (0x00FFFFFF), B(0x0000FFFF) or
C(0x000000FF).
-257 File name error; Filename cannot contain any directory information:
{name}
This error occurs when a filename with any directory information is
passed to the SCPI command RADio:ARB:GENerate:SINE. The parameter
is the bad filename.
-257 File name error; The filename specified for saving or recalling is not
valid. Note that only the STATE memory subsystem can be used with
save/recall. Name supplied: {name}
Save/Recall can only operate on the STATE memory subsystem. The
parameter is the bad filename.
-257 File name error; Invalid Filename
An invalid filename was encountered. Please refer to the user
manual for guidelines on filename format.
-257 File name error; No file has been selected for use as the PRAM file.
Please select an existing file before trying to use PRAM.
No file has been selected for use as the PRAM file. Please select
an existing file before trying to use PRAM.
-257 File name error; Waveform file not selected.
Waveform file or sequence file must be selected when turning ARB
on.
-257 File name error; {name}
There was a problem with the specified file name. Check the syntax
and try again.
-256 File name not found; User preset file could not be found: a user
preset state must be saved before the preset type can be set to User.
The instrument could not find a user-saved preset state. The user
must first save a preset state before the Preset Type can be set to
User.
-256 File name not found; {name}
The file specified does not exist.