User manual

lpErrorDoc.txt
-271 Macro syntax error
A syntactically legal macro program data sequence, written
in accordance with IEEE 488.2, 10.7.2, could not be executed
due to a syntax error within the macro definition
(see IEEE 488.2, 10.7.6.3).
-270 Macro error
A macro-related execution error occurred. This error message
is used when the device cannot detect the more specific
errors described for errors -271 through -279.
-261 Math error in expression
A syntactically legal expression program data element could
not be executed due to a math error. For example, a
divide-by-zero was attempted. The definition of a math
error is device-specific.
-260 Expression error
An expression data element-related error occurred. This
error message is used when the device cannot detect the
more specific errors described for errors -261 through -269.
-259 LAN setup error; Default gateway IP address is inconsistent with
subnet mask or IP class rules. Offending entry is changed.
The default gateway IP address is invalid. Check the entry for
syntax and numerical errors and try again. Verify that the subnet
mask is consistent with this address (host bits should not be all
0s or 1s). If no subnet is specified, verify that the address is
valid with respect to the default class netmasks: A (0x00FFFFFF),
B(0x0000FFFF) or C(0x000000FF).
-259 LAN setup error; The instrument could not extend lease for a dynamic
IP address using DHCP. The LAN interface will be shut down until a new
address can be obtained again.
An initial DHCP LAN configuration was obtained. However, when the
lease expired, the instrument was unable to extend it and as a
result lost the previously assigned IP address. Refer to the
internet DHCP protocol documentation for further information.
-259 LAN setup error; The instrument could not obtain a LAN configuration
using DHCP. The LAN interface is not initialized. The instrument will
continue to try to obtain a LAN configuration.
The instrument was configured to use DHCP to obtain the LAN
interface configuration. The instrument was not able to negotiate
with surrounding DHCP servers. Either DHCP servers are not present,
or they are not allowing this device to obtain an IP address.
Possibly, the LAN cable is disconnected or is broken. The
instrument will try to contact DHCP servers at regular time
intervals. The LAN setup menu will be updated if LAN configuration
is obtained.
-259 LAN setup error; The instrument LAN setting could not be changed.
An initial LAN configuration was obtained and cannot be changed.
Page 11