150 Chpater 6
Custom Arb Waveform Generator
Working with Filters
To Create a User-Defined FIR Filter with the FIR Values Editor
In this procedure, you use the FIR Values editor to create and store an 8- symbol, windowed, sinc 
function filter with an oversample ratio of 4. The Oversample Ratio (OSR) is the number of filter 
coefficients per symbol.
You can define from 1 to 32 FIR coefficients, where the maximum combination of symbols and 
oversample ratio is 1024 coefficients.
The FIR Values editor allows a maximum filter length of 1024 coefficients, but the PSG hardware is 
limited to 512 symbols for arbitrary waveform generation, and 64 symbols for real- time waveform 
generation. The number of symbols equals the number of coefficients divided by the oversample ratio. 
If you enter more than the maximum number of symbols, the PSG cannot use the filter; it decimates 
the filter (throws away coefficients) until the required condition is met and then uses the filter, but 
fine resolution may be lost from the impulse response.
1. Press 
Preset.
2. Press 
Mode > Custom > Arb Waveform Generator > Digital Mod Define >  Filter
or
Mode > Custom > Real Time I/Q Baseband > Filter
3. Press  Define User FIR > More (1 of 2). 
4. Press 
Delete All Rows >  Confirm Delete Of All Rows > More (2 of 2). 
This brings up the FIR Values editor and clears the table of existing values.
5. Press 
Edit Item.The  Value field for coefficient 0 should be highlighted.
6. Use the numeric keypad to type the first value (−0.000076) from the following table and press 
Enter. As you press the numeric keys, the numbers are displayed in the active entry area. (If you 
make a mistake, you can correct it using the backspace key.) Continue entering the coefficient 
values from the table until all 16 values have been entered.