Installation guide

24 Chapter 3
Operation Verification
Checking the Output Power
NOTE Limit values are due to power measurement uncertainty.
Problems with Output Power
If a problem occurs when verifying output power levels, perform the following procedure:
1. Verify the power sensor being used in this test is the appropriate sensor.
2. Verify the power sensor calibration factors, if required, are correct.
3. Verify the power sensor is properly calibrated to the power meter.
If the problem continues, perform a Power Level Accuracy performance test. Refer to the
Troubleshooting chapter of the E8257D/67D PSG Signal Generators Service Guide.
E8267D
Option 532
and
Option 544
250 kHz 12 dBm / ±1.0 N/A N/A N/A
3.2 GHz 12 dBm / ±1.0 N/A N/A N/A
20 GHz 14 dBm / ±1.0 N/A N/A N/A
40 GHz 12 dBm / ±1.0 N/A N/A N/A
44 GHz 10 dBm / ±0.9 N/A N/A N/A
Table 3-6 Frequency and Power Level Limits (For PSGs with s/n prefixes >US4928/MY4928)
Instrument
Model
Frequency Amplitude / Limits
Standard Option 1EA Option 1E1 Options 1EA
and 1E1