Installation guide
16 Chapter 3
Operation Verification
Checking the Maximum Leveled Power
Table 3-1 Frequency and Power Level Limits (For PSGs with s/n prefixes <US4928/MY4928)
Instrument
Model
Amplitude Frequency
# Points
Standard Option
1EA
a
a.Option 1EH or and Option UNW can degrade Option 1EA’s maximum power specifications. Refer to the Data Sheet.
Option
1E1
Options
1EA and
1E1
Start Stop
E8257D
Option 520
13 dBm 16 dBm 11 dBm 15 dBm 250 kHz 3.2 GHz 50
13 dBm 20 dBm 11 dBm 18 dBm 3.21 GHz 20 GHz 200
E8257D
Option 532
9 dBm 15 dBm 7 dBm 14 dBm 250 kHz 3.2 GHz 50
9 dBm 18 dBm 7 dBm 16 dBm 3.21 GHz 20 GHz 200
9 dBm 14 dBm 7 dBm 12 dBm 20.1 GHz 31.8 GHz 200
E8257D
Option 540
9 dBm 15 dBm 7 dBm 14 dBm 250 kHz 3.2 GHz 50
9 dBm 18 dBm 7 dBm 16 dBm 3.21 GHz 20 GHz 200
9 dBm 14 dBm 7 dBm 12 dBm 20.1 GHz 40 GHz 200
E8257D
Option 550
and
Option 567
5 dBm 14 dBm 3 dBm 13 dBm 250 kHz 3.2 GHz 50
5 dBm 14 dBm 3 dBm 13 dBm 3.21 GHz 20 GHz 200
5 dBm 11 dBm 3 dBm 9 dBm 20.1 kHz 65 GHz 200
5 dBm 10 dBm 3 dBm 8 dBm 65.1 GHz 67 GHz 100
E8267D
Option 520
13 dBm N/A N/A N/A 250 kHz 3.2 GHz 50
18 dBm N/A N/A N/A 3.21 GHz 20 GHz 200
E8267D
Option 532
and
Option 544
12 dBm N/A N/A N/A 250 kHz 3.2 GHz 50
14 dBm N/A N/A N/A 3.21 kHz 31.8 GHz 200
12 dBm N/A N/A N/A 32.1 GHz 40 GHz 200
10 dBm N/A N/A N/A 40.1 GHz 44 GHz 100
E8663B 15 dBm N/A N/A N/A 250 kHz 3.2 GHz 50
18 dBm N/A N/A N/A 3.21 GHz 9 GHz 100