Installation guide
Chapter 3 15
Operation Verification
Checking the Maximum Leveled Power
10. Press Return > Sweep > Freq. The SWEEP annunciator is now on and a continuous sweep will begin.
During the sweep, the progress bar becomes active on the display, indicating the status of the
current sweep.
11. Watch the display and ensure that the Unlevel annunciator does not appear at any time during
a full sweep.
The Unlevel annunciator may turn on and off, or remain on; both conditions indicate a failure.
If this occurs, refer to “Maximum Leveled Power Check Failure” on page 18 for further
instructions.
12. Press
Sweep > Off.
13. Repeat steps 5 through 12 for each frequency range specific to your instrument model number
and options.