Agilent Technologies E8247C and E8257C/D Option H30 User’s and Service Guide Supplement
Agilent Technologies E8247C and E8257C/D Option H30 User’s and Service Guide Supplement Use this manual with the following documents: PSG Family User’s Guides: E8251-90253 E8251-90259 E8251-90353 E8251-90359 Manufacturing Part Number: E8251-90072 Printed in USA November 2005 © Copyright 2003, 2005 Agilent Technologies, Inc. All rights reserved.
Warranty Statement THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Safety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document. WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met. CAUTION Caution denotes a hazard.
iv
Contents E8247C and E8257C\D Option H30 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Nominal Performance Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Operation and Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Documentation Conventions . . . . . . . . . . . . . . . . . . . . . . . .
Contents Contents-2
E8247C and E8257C\D Option H30 User’s and Service Guide Supplement 1
Description The Option H30 adds a mixer and coax switch to enable the E8247C or the E8257C/D to upconvert a RF signal (input is provided on the rear panel) to a higher frequency which will be the sum and difference of the RF input signal and frequency set on the E8247C or E8257C/D. The coaxial switch allows the E8247C or E8257C/D with Option H30 to be switched from standard synthesizer mode to upconverter mode. This allows the E8247C to be used as a standard source instrument, as well as an upconverter.
Nominal Performance Characteristics Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.
Figure 1 Upconverted RF Flatness 20 15 IF Input frequency vs. flatness over 1 GHz (dB) for fundimental mixing 10 5 0 0. 4 0. 8 1 1. 4 1. 6 2 2. 4 2. 6 3 3. 2 3. 6 3. 8 4 4. 4 4. 8 5 5. 2 5. 6 5.
Operation and Performance Verification Documentation Conventions The following key conventions are used throughout this document. • [HARDKEYS] are labeled front panel keys. • SOFTKEYS are unlabeled keys whose function is indicated on the instrument display. Option H30 Operation (520 or 540) To use the E8247C and E8257C/D Option H30 as a standard instrument, refer to the standard User’s Guide.
Performance Verification The E8247C and E8257C/D Option H30 functionality can be verified using the following equipment and methods: The E8247C and E8257C/D Option H30 must first meet standard product specifications with the exception that the power output may be degraded by up to 2.0 dBm. (Subtract 2.0 dBm from the Option 1E1 output power specification).
Troubleshooting If there is no signal a problem has occurred. First check the equipment setup, verifying outputs on the sources and connections. Refer to Figure 4 through Figure 7 . Verify that the E8247C and E8257C/D Option H30 Upconverter switch is working: MID Board or Coaxial Switch The following procedure will aid in determining whether the MID board (E8251-60089) or the coax switch (87222E) is functional. Step 1. Select upconverter mode on/off, then listen for the coax switch to click.
Upconverter Mixer The following procedure will aid in determining if the upconverter mixer is functional: CAUTION Possible damage may occur if RF upconverter input levels exceed 20 dBm or if reverse power is applied at the RF output connector. First verify all cable configurations are correct before applying a signal to the instrument. The RF step attenuator cable to the coax switch is functional if the instrument operates as a standard synthesizer.
Repair Information Figure 4 Top View W105 W102 Figure 5 W104 A9 MID Board W101 W103 Semi-rigid Cables (front view) MP101 W102 A100 Mixer W103 W105 SW1 User’s and Service Guide Supplement W104 W101 9
Figure 6 Ribbon Cable A100 MP101 W106 A9 Figure 7 Bracket (rear panel) Washer Nut MP108 MP107 10 User’s and Service Guide Supplement
Replaceable Parts Description Agilent Part Number A100 Mixer 6 GHz to 46 GHz (Opt H30 with Opt 540) 0955-1447 1 A100 Mixer 6 GHz to 26 GHz (Opt H30 with Opt 520) 0955-0488 1 SW1 Coax transfer switch, 4 port, 50 Ω, 50 GHz, 87222E 87222-60015 1 W101 RF Cable Coupler- SW (Opt H30 with Opt 540) E8251-20082 1 W101 RF Cable Coupler- Switch (Opt H30 with Opt 520) E8251-20173 1 W102 RF Cable Attenuator - SW E8251-20083 1 W103 RF Cable Mixer IF - Rear Panel E8251-20084 1 W104 RF Cable
Safety and Regulatory Information Introduction Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition. Connector Care and Cleaning If alcohol is used to clean the connectors, the power cord to the instrument must be removed.
Compliance with German Noise Requirements This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3. GSGV Deutschland). Acoustic Noise Emission/Geraeuschemission LpA<70 dB Lpa<70 dB Operator Position am Arbeitsplatz Normal Operation normaler Betrieb per ISO 7779 nach DIN 45635 t.
Warnings WARNING The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, which if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Warnings applicable to this instrument are: WARNING If this instrument is not used as specified, the protection provided by the equipment could be impaired.
Cautions CAUTION The CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like, which if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. Cautions applicable to this instrument are: CAUTION Always use the three-prong ac power cord supplied with this instrument.
Instrument Markings ! When you see this symbol on your instrument, you should refer to the instrument’s instruction manual for important information. This symbol indicates hazardous voltages. The laser radiation symbol is marked on products that have a laser output. This symbol indicates that the instrument requires alternating current (ac) input. The CE mark is a registered trademark of the European Community. If it is accompanied by a year, it indicates the year the design was proven.
Contacting Agilent By internet, phone, or fax, get assistance with all your test and measurement needs. This information supersedes all prior contact information. Online assistance: www.agilent.
18 User’s and Service Guide Supplement