Troubleshooting
Troubleshooting Assembly–Level Problems
705 F/2 Test
1. Using an oscilloscope, check the 10 MHz Synthesis line on P33–15. The signal period should be
100 ns with an amplitude of approximately 3V
pp
.
If the signal is good, replace the A6 Frac–N.
2. If the signal is bad, turn the signal generator power off, remove the A6 Frac–N, and measure
P33–15 again.
If the signal is good, replace the A6 Frac–N.
3. If the signal is still bad, measure P41–5 on the A7 Reference.
If the signal is good at P41–5, replace the A31 Motherboard.
If the signal is bad at P41–5, replace the A7 Reference.
706 FM Path Test
1. Using an oscilloscope, check P31–22 (FM_On_H):
a. Loop self–test 706
b. Check for a TTL transition from 0 Vdc to +5 Vdc.
If the signal is not present, refer to A5 Sampler troubleshooting.
2. If the signal is present, check P31–11 (FM):
a. Loop self–test 706.
b. Check for a 0 to 3.3 Vdc transition.
If the signals is present, replace the A6 Frac–N.
If the signal is not present, refer to A11 Pulse/Analog Modulation Generator
troubleshooting.
707 Sweep Phase Detector Test
1. Replace the A6 Frac–N.