Specifications

1-12
Troubleshooting
Troubleshooting Assembly–Level Problems
Symptom: Dark display
1. Ensure that the front panel green power on LED is on. If not, refer to “Power Supply
Troubleshooting” on page 1-17.
2. Press any front panel key to disable the screen saver.
3. If the display is still bad, check P12 on the A31 Motherboard for the following signals:
If the 5.2 Vdc line is bad, refer to “Power Supply Troubleshooting” on page 1-17.
If either the LIGHT_EN or VDISP are bad, check them at P22116 and P22155 respectively
on the A31 Motherboard. If either signal is bad at P221, replace the A18 CPU.
If all the above signals are good, check the output of the dc to ac inverter as follows:
a. Remove the front panel from the frame, and lay the front panel face down.
b. With CN21 open, there is typically a 160 Vrms 40 kHz signal. If not, replace the dc to ac
inverter.
If the signal is there, replace the LCD.
Signal State
P121 5.2 Vdc
P123 LIGHT_EN >3 Vdc
P124 VDISP Brightness = 50, VDSP < 50 mV
Brightness = 1, VDSP = 3 Vdc