Specifications

2-38
Assembly Replacement
A18 CPU
A18 CPU
Tools Required
T-10 driver
Removal Procedure
Refer to Figure 2-17 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the front panel facing you.
4. Using the T-10 driver, remove the four screws from the Baseband Generator cover
(E8267C Only).
5. Disconnect the following ribbon cables:
W19 from A18J2
W66from A18P2
W18 from A18J1
W64 from A14J13
W90 from A14J21
6. Lift the retention levers and disconnect the A18 CPU from the A31 Motherboard.
7. Lift the A18 CPU out of the slot.
Replacement Procedure
Reverse the order of the removal procedure.
Send command "DIAG:FILE:REST", this copies data to the CPU board.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.