Specifications
5
1. Press [Preset].
2. Press [Frequency] > [10] > {GHz}.
3. Press [Amplitude] > [0] > {dBm}.
4. Press [RF On/Off].
5. Press {Leveling Mode} > {External Detector}. This configuration deactivates the internal ALC
detector and switches the ALC input path to the
front panel ALC INPUT connector. The EXT
indicator is activated in the AMPLITUDE area
of the display.
NOTE For signal generators with Option 1E1, notice that
the ATTN HOLD (attenuator hold) enunciator is
displayed. During external leveling, the signal
generator automatically uncouples the attenuator
from the ALC system for all external leveling points.
While in this mode, RF output amplitude
adjustment is limited to –20 to +25 dBm, the
adjustment range of the ALC circuitry. For more
information, see External Leveling with Option 1E1
Signal Generators on page 6.
6. Observe the coupling factor printed on the
directional coupler at the detector port.
Typically, this value is -10 to -20 dB.
Enter the positive dB value of this coupling factor
into the signal generator.
7. Press {More (1 of 2)} > Leveled output power is now available at
{Ext Detector Coupling Factor} > [16] (or the the output of the directional coupler.
positive representation of the value listed at the
detector port of the directional coupler) > {dB}.
NOTE While operating in external leveling mode, the
signal generator’s displayed RF output amplitude is
affected by the coupling factor value, resulting in a
calculated approximation of the actual
RF output amplitude.
To determine the actual RF output amplitude at the
point of detection, measure the voltage at the
external detector output and refer to Figure 3-2.
Configure the signal generator