Specifications

19
1. Press {Load/Store}.
2. Press {Store to File}.
3. Enter the file name FLATCAL2 using the
alphanumeric soft keys and the numeric keypad.
4. Press {Enter}. The user flatness correction array file FLATCAL2 is
now stored in the memory catalog as a UFLT file.
Save the User Flatness Correction Data to the
Memory Catalog
This process allows you to save the
user flatness correction data as a
file in the signal generator’s memory
catalog. With several user flatness
correction files saved to the memory
catalog, specific files can be
recalled, loaded into the correction
array, and applied to the RF output
to satisfy various RF output flatness
requirements.
Applying the user flatness
correction array
1. Press [Return] > [Return] > {Flatness Off On}. This applies the user flatness correction array to
the RF output. The UF indicator is activated in the
AMPLITUDE section of the signal generator’s
display and the frequency correction data
contained in the correction array is applied to the
RF output amplitude of the mm-wave
source module.
1. Press [Preset].
2. Press [Amplitude] > {More (1 of 2)} >
{User Flatness} > {Configure Cal Array} >
{More (1 of 2)} > {Preset List} > {Confirm Preset}.
3. Press [More (2 of 2)] > {Load/Store}.
4. Ensure that the file FLATCAL2 is highlighted.
5. Press {Load From Selected File} > This populates the user flatness correction array
{Confirm Load From File}. with the data contained in the file FLATCAL2. The
user flatness correction array title displays User
Flatness: FLATCAL2.
6. Press [Return] > {Flatness Off On}. This activates flatness correction using the data
contained in the file FLATCAL2.
Recalling and applying a user
flatness correction array
Before performing the steps in this
section, complete the section
Creating a User Flatness Correction
Array Using a MM-Wave Source
Module.