Specifications
17
Perform the user flatness correction
NOTE If you are not using an
Agilent E4416A/17A/18A/19A power
meter, or if you do not have GPIB
interface capability, you can perform
the User Flatness Correction
manually. For instructions, see
“Performing the User Flatness
Correction Manually.”
1. Press {More (1 of 2)} > {User Flatness} > This performs the user flatness correction. The
{Do Cal}. signal generator enters the user flatness correction
routine and a progress bar is shown on the display.
2. When prompted, press {Done}. This loads the amplitude correction values into the
user flatness correction array.
If desired, press {Configure Cal Array}. This opens the user flatness correction array,
where you can view the list of defined frequencies
and their calculated amplitude correction values.
The user flatness correction array title displays
User Flatness: (UNSTORED) indicating that the
current user flatness correction array data has not
been saved to the memory catalog.