Specifications

10
1. Zero and calibrate the power
meter and power sensor.
NOTE If you are using a power meter/sensor combination
other than the Agilent E4416A and E4413A, it may
be necessary to:
a. Enter the appropriate power sensor calibration
factors into the power meter.
b. Enable the power meters cal factor array.
For operating information on your particular power
meter/sensor, refer to their operating guides.
1. Connect the equipment as shown in Figure 3-5.
NOTE During the user flatness correction process, the
power meter is slaved to the signal generator via
GPIB. No other controllers are allowed on the
GPIB interface.
Configure the power meter
Figure 3-5.
User Flatness
Correction Equipment
Setup.
Connect the equipment
LINEOI
INTERNAL PULSE GENERATOR
GATE/PULSE
TRIGGER INPUT
50
VIDEO
OUT
50
SYNC
OUT
50
RF OUTPUT
50
REVERSE PWR
1W MAX 0V DC
!
LP OUTPUT
ALC INPUT
EXT 2 INPUT
EXT 1 INPUT
Sweep/
List
AM
FM/
ΦM
LF Out
Pulse
Utility
MENUS
Save Recall Trigger Help
Mod
On/Off
RF
On/Off
7 8 9
4 5 6
1 2 3
0 .
+/-
Hold
Incr
Set
E8254A
250 kHz 40 GHz
PSG -A Series Signal Generator
Frequency
Amplitude
ReturnLocalPrint
FREQUENCY
AMPLITUDE
40.000 000 000 000
GHZ
10.00
dBm
MOD
ON
RF
ON
RF output
Signal
generator
Input
port
Power
meter
Cables
and other
devices
ALC input
Source
module
GPIB
Flatness
corrected
output port
In
Out
Device
under test
Power
sensor