User`s guide
Agilent PXT Wireless Communications Test Set
User’s Guide
66
The message is not sent during a period of time when the UE is not listening to the DL (Connected Mode
DRX or Measurement Gap). The TA command is still sent, but at the next available opportunity (for
example: when the onDuration occurs). If the TA command is a periodic one (see Periodic Timing Advance
Frequency) and several are sent when the UE is in DRX, only one will be sent when the next onDuration is
available.
Mode
BSE
Initial S/W Revision
6.5
Key Path
Mode > BSE > Func > More > Timing Advance
Timing Advance Value
Enter the value you wish to send to the UE via the Send Message key described above. This value is used
for custom or periodic Timing Advance commands.
Mode
BSE
Range
0 to 63
Preset
31 (Indicates no change)
Initial S/W Revision
6.5
Key Path
Mode > BSE > Func > More > Timing Advance
Periodic Timing Advance Frequency
Sends a timing advance command to the UE with the frequency (every < x > milliseconds) specified here.
Mode
BSE
Range
0 to 10240 ms
Preset
0 (disables feature)
Dependencies and/or Couplings
When set to 0, Timing Advance Expiry Timer is not available.
Initial S/W Revision
6.5
Key Path
Mode > BSE > Func > More > Timing Advance
Timing Advance Expiry Timer
This setting is the number of milliseconds the PXT continues to send periodic timing advance commands
to the UE, while the UE is not transferring any data on an SRB or DRB, before it stops doing so.
Once the DL or UL data begins transferring again, the periodic timing advance commands restart.
Mode
BSE
Range
0 to 10240 ms
Preset
0 (disables feature)
Dependencies and/or Couplings
Grayed out if periodic timing advance frequency is set to 0.
Initial S/W Revision
6.5
Key Path
Mode > BSE > Func > More > Timing Advance