User`s guide

Agilent PXT Wireless Communications Test Set
User’s Guide
65
Median CQI Start
This setting starts/initiates the collection of CQI reports from UE.
For aperiodic CQI reports this setting requires the PXT to start asking for aperiodic CQI reports
implicitly.
For periodic CQI reports the UE will continue to send CQI reports periodically; this setting will
start collect periodic CQI reports to generate the medium CQI statistics.
For this to work properly, the scenario must contain the appropriate CQI Report
Configuration (either periodic or aperiodic) in the RRC Setup message information.
This enables the UE to generate the correct CQI reports.
Mode
BSE, FDD
Initial S/W Revision
6.3
Key Path
Mode > BSE > Func > More > CQI Median
Notes The calculation for the Median CQI stops automatically when the
desired number of CQI reports are obtained. Alternatively, you can stop
the calculation by manually pressing Mode > BSE > Func > More >
CQI Median > Mediam CQI Stop.
Median CQI Stop
Enables you to manually stop the calculation for the Median CQI.
Mode
BSE, FDD
Initial S/W Revision
6.3
Key Path
Mode > BSE > Func > More > CQI Median
Notes The calculation for the Median CQI stops automatically when the
desired number of CQI reports are obtained. Alternatively, you can stop
the calculation by manually pressing this key.
Timing Advance
This key accesses the menu that enables you to test the ability of the UE to respond to timing advance
commands while the UE is in RRC Connected mode. The following keys are available:
Mode
BSE
Initial S/W Revision
6.5
Key Path
Mode > BSE > Func > More
Send Message
Sends a Timing Advance (TA) MAC Control Element containing the value held in the menu key below this
one. If the UE is RRC Connected, but no data is being sent, the Timing Advance MAC Control Element is
sent inside its own transport block. Otherwise, the MAC Control Element is added to a transport block that
is carrying data.
Pressing this menu key also starts the sending of Periodic Timing Advance commands with the frequency
specified, if the Periodic Timing Advance Frequency
setting below is set to something other than 0.