User`s guide

Agilent PXT Wireless Communications Test Set
User’s Guide
63
Mode
BSE, FDD
Range
1 to 1000000
Preset
12913
Initial S/W Revision
6.3
Key Path
Mode > BSE > Func > More > RCT > PM-dsg
Confidence Level
Setting this menu key to On enables the PXT to determine the probability of whether the PM-dsg
measurement will pass or fail, based on the table in 3GPP TS 36.521-1 Appendix G.4.4-1. This may cause
the Pm-dsg measurement to stop before the Max Frame Size
has been reached.
Setting this menu key to Off enables the Pm-dsg measurement to run for the number of transport blocks
set using the Max Frame Size
menu key.
Mode
BSE, FDD
Range
On | Off
Preset
Off
Initial S/W Revision
6.3
Key Path
Mode > BSE > Func > More > RCT > PM-dsg
Test Start
This menu key starts the Pm-dsg measurement.
Mode
BSE, FDD
Initial S/W Revision
6.3
Key Path
Mode > BSE > Func > More > RCT > PM-dsg
Test Stop
This menu key stops the Pm-dsg measurement.
Mode
BSE, FDD
Initial S/W Revision
6.3
Key Path
Mode > BSE > Func > More > RCT > PM-dsg
Statistical CQI Performance
The Statistical CQI Performance measurement is used as part of an RCT system to perform test cases in
36.521-1, section 9. As described in this specification, the measurement sends transport blocks containing
fixed MAC padding data to the UE and counts the ACKs and NACKs sent in response. The ratio of
(NACK)/(NACK+ACK) determines the block error rate.
Mode
BSE, FDD
Initial S/W Revision
6.3
Key Path
Mode > BSE > Func > More > RCT
Max Frame Size
This setting determines the maximum number of transport blocks sent when Test Start is selected.
Mode
BSE, FDD
Range
1 to 1000000