User`s guide
Agilent PXT Wireless Communications Test Set
User’s Guide
60
RCT
Accesses the following parameter menus:
Statistical Throughput
Pm-an
Pm-dsg
Statistical CQI Performance
Key Path: Mode > BSE > Func > More
Statistical Throughput
The Statistical Throughput measurement enables testing of 36.521-1, section 7 test cases. (Section 7 test
cases are all SISO.)
The Statistical Testing of Receiver Characteristics, defined in 3GPP TS 36.521-1 Appendix G.2 describes this
measurement.
This menu key accesses the menu enabling you to control Max Frame Size, Confidence Level, Test Start,
and Test Stop and is only available in FDD.
Mode
BSE, FDD
Initial S/W Revision
6.4
Key Path
Mode > BSE >Func > More > RCT
Max Frame Size
This setting determines the maximum number of transport blocks sent when Mode > BSE > Func > More
> RCT > Statistical Throughput > Test Start is selected.
If Confidence Level is set to On, the test may stop before reaching the maximum
number of frames.
Mode
BSE, FDD
Range
1 to 1000000
Preset
2466
Initial S/W Revision
6.4
Key Path
Mode > BSE > Func > More > RCT > Statistical Throughput
Confidence Level
Setting this menu key to On enables the PXT to determine the probability of whether the Statistical
Throughput measurement will pass or fail, based on the table in 3GPP TS 36.521-1 Appendix G.2. This may
cause the statistical throughput measurement to stop before the Max Frame Size
has been reached.
Setting this menu key to Off enables the statistical throughput measurement to run for the number of
transmissions set using the Max Frame Size
menu key.
Mode
BSE, FDD
Range
On/Off
Preset
Off