User`s guide

Agilent PXT Wireless Communications Test Set
User’s Guide
52
Handover Setup Mode
Enables you to set parameters for performing a blind handover.
When set to Blind
, after sending the selected handover message, the PXT automatically adjusts the DL
Frequency, UL Frequency, and Physical Cell Id to the values specified below. This makes it possible to
perform a handover with only one cell present at a time (without option BB1).
Mode
BSE, FDD
Range
Default | Blind
Preset
Default
Initial S/W Revision
6.4
Key Path
Mode > BSE > Func > Handover
Destination DL-Carrier Freq
Sets the destination downlink carrier frequency for a blind handover.
Mode
BSE
Range
0 to 65535
Preset
300
Dependencies and/or Couplings
Handover Setup Mode must be set to ‘Blind’.
Initial S/W Revision
6.4
Key Path
Mode >BSE > Func > Handover
Destination UL-Carrier Freq
Sets the destination uplink carrier frequency for blind handover.
Mode
BSE
Range
0 to 65535
Preset
18300
Dependencies and/or Couplings
Handover Setup Mode must be set to ‘Blind’.
Initial S/W Revision
6.4
Key Path
Mode >BSE > Func > Handover
Target Phys Cell Id
Sets the target cell ID for the Blind Handover.
Mode
BSE
Range
0 to 503
Preset
0
Dependencies and/or Couplings
Handover Setup Mode must be set to ‘Blind’.
Initial S/W Revision
6.4
Key Path
Mode >BSE > Func > Handover