User`s guide
Agilent PXT Wireless Communications Test Set
User’s Guide
32
This setting is primarily used with testing MIMO configurations.
Maximum RF2 Output power is -10dBm.
Mode
BSE, SA
Range
–100 dB to +100 dB
Units
dB
Preset
0
Initial S/W Revision
6.0
Key Path
Mode > BSE/SA > Config > Amplitude Offsets
RF1 Input Power Offset
Enables you to enter a value to adjust the measurement results.
Enter a positive value to correct for a loss. For example: For 20dB of loss, enter +20dB.
This loss is applied globally. Frequency dependent definition of loss is available by using The Tabular
Method described above.
This setting does NOT adjust the input attenuation. You must set the
attenuation by setting Atten > Attenuation to Manual and then setting Atten >
RF1 Input Attenuation to the desired level.
Mode
BSE, SA
Range
–100 dB to +100 dB
Units
dB
Preset
0
Initial S/W Revision
6.0
Key Path
Mode > BSE/SA > Config > Amplitude Offsets
RF2 Input Power Offset
This function is currently not available.
General Config
Accesses configuration settings that apply to all types of measurements.
Mode
BSE, SA
Initial S/W Revision
6.4
Key Path
Mode > BSE/SA > Config