User`s guide
Agilent PXT Wireless Communications Test Set
User’s Guide
200
Channel Power Measurement (FDD/TDD)
Key Path:
Mode > SA > LTE > Channel Power
Figure 6-5: Channel Power measurement
The Channel Power measurement captures 10 sub-frames as represented by the indices 0-9 in the Figure
above. Each of the ten intervals can be edited to a specific start time and duration time for general
purposes. For example, you might want to exclude 20 µsec transition periods on the sub-frame index 0
boundaries. You would then edit index 0 by changing the range from 0 − 1000 µsec to 20 − 980 µsec. This
is useful in situations where you wish to exclude transient periods from the measurement period interval.