User`s guide
Agilent PXT Wireless Communications Test Set
User’s Guide
146
Setup and Operation
General Call Setup Procedure
1. Press Emulator Mode > BSE > Mode Setup > Call Scenario to select the call scenario you wish to
use.
All instrument parameters must match the selected scenario file in order to
successfully connect to the UE.
Figure 3-3: Example Call Scenario List
2. Adjust Cell ID, PHICH Resource, CFI to align with the call scenario selected by pressing Mode > BSE
> Mode Setup > More > PHY Settings. Refer to the section in this manual entitled, PHY Settings on
page 88
for more information.
3. To specify security settings, select Mode > BSE > Mode Setup > More > Security. You can enter the
algorithm used as well as the values for K, OP, OPc, and AMF. Refer to the section in this manual
entitled, Security on page 130
for more information.
4. Select Mode > BSE, and set the Emulator to Run. If it is already running, select Stop and reset it to
Run, again. The UE is now ready to be connected.
5. When the UE tries to connect, the transmitted messages are shown on the Message Display Window.
If the connection is successful, the UE Status Information changes to “CON” (connected).