User`s guide
Agilent PXT Wireless Communications Test Set
User’s Guide
135
Security – Key Menu 2
Replay
When selecting On, the PXT automatically populates the Information Element (within the
SECURITY_MODE_COMMAND) called, “Replayed UE Security Capabilities” with the exact capability
content received by the UE within the previously received ATTACH_REQUEST UL EPS Mobility
Management (EMM) Message.
When selecting Off, you manually define this Information Element within the
SECURITY_MODE_COMMAND of the currently loaded scenario file.
The SECURITY_MODE_COMMAND is a DL EMM Message, sent to the UE
during connection setup with security enabled. It enables ciphering and
integrity at NAS.
Mode
BSE
Range
ON|OFF|1|0
Preset
ON|1
State Saved
No. Instrument always boots up in the “On” state.
Initial S/W Revision
6.1
Key Path
BSE > Mode Setup> More > Security > More
Preset
Returns all settings to their default factory values.
Key Path: Front-panel key
Print
Function is not currently available.
Key Path: Front-panel key
Recall
Enables you to load parameter settings from the state files saved in registers. See Save State menu for more
information.
Key Path: Front-panel key
Recall State
Enables you to recall state files from registers and delete state files.
Key Path: Recall