Technical data

12 Getting Started Guide
1 Quick-Start
Shipping problems?
If the shipping materials are damaged or the contents of the container are incomplete:
Contact the nearest Agilent Technologies office.
Keep the shipping materials for the carriers inspection.
If you must return an test set to Agilent Technologies, use the original (or
comparable) shipping materials. See “Returning a test set for Service" on page 111.
Test Set Location and Mounting Requirements
Locating the test set
Make sure that the fan inlet and exhaust vent areas on the sides, bottom, and back of the
test set chassis are not obstructed. The minimal required clearance is 2 inches. Airflow
restrictions cause additional airflow noise and cause the fans to speed up so they can
draw in enough air for the required cooling resulting in excessive audible noise.
Cooling and rack mounting
Do not mount the test set side-by-side with any other instrument with side-by-side
ventilation. Make sure the exhaust air from one instrument is directed away from the
inlet of any other instrument (heated air directed into the inlet of an instrument can cause
instrument failure due to excessive operating temperatures). The test set draws air in
from the sides and bottom of the chassis and exhausts air from the back.