Technical data
Chapter 6 81
Front Panel Sequence Setup
Front Panel List Sequencer Setup
Front Panel Sequence Setup
Basic Measurement setup
If you are including any of the basic measurements in the sequence, set the parameters from the front
panel from within the Sequence Analyzer mode as follows:
Step 1. Press
Meas Setup, Basic Meas Setup and select from the following:
•
Basic Transmit Power Setup
• Freq and Phase Error Setup
• Basic Discrete PAvT Setup
• Basic IQ Data Setup
Step 2. If you have selected Basic Transmit Power, then select a radio standard (Meas Setup,
Acquisition Setup, Radio Setup,
Radio Standard. (The choices are None, GSM, EDGE,
WCDMA, cdma2000, 1xEV-DO, LTE, LTE-TDD, or TDSCDMA.) Optionally, set
Upper and
Lower limits to the received power (in dB or -dB). (When you select Radio Standard, further
selections include
Digital IF BW, Filter Type, Filter BW, and Filter Alpha settings.)
Step 3. If you have selected
Freq and Phase Error Setup, specify the Digital IF BW, in MHz, and set
the
Limit: FreqError (On or Off, plus the limit value in ppm).
Step 4. If you have selected
Basic Discrete PAvT Setup, specify the Filter Type (Wide, Narrow,
Very Narrow, or Extremely Narrow).
Step 5. If you have selected
Basic IQ Data Setup, then you need to specify the Digital IF BW, in
MHz, and specify the
Meas Type (Raw IQ Data or Amplitude and Phase).
Mode-specific measurement setup
If you are running measurements other than the basic measurements, then you have to change modes to
set up the associated measurement parameters from the front panel. For more help in each mode, use the
mode online help which can be accessed by pressing the Help hardkey from within each mode:
Step 1. Press the
Mode hardkey.
Step 2. Press the softkey for one of the modes which run within the Sequence Analyzer (
GSM, EDGE,
WCDMA, cdma2000, 1XEVDO, LTE, LTE-TDD, or TDSCDMA
) to select the licensed mode.
Step 3. In the mode, use the
Meas Setup hardkey to set up the measurement parameters for the
measurements you are performing.
Acquisition setup
If you have changed modes to set up the measurements, change back to Sequence Analyzer mode. To set
up an acquisition, proceed as follows:
Step 1. Press
Meas Setup, Acquisition Setup.
Step 2. Use the
Number of Acquisitions softkey to set the total number of acquisitions in the
sequence. Alternatively, add or remove acquisitions using the
Insert Before Acquisition and
Delete Acquisition softkeys.