Technical data

80 Chapter 6
Front Panel Sequence Setup
Front Panel List Sequencer Setup
Front Panel Sequence Setup
List sequencer global parameter setup
The list sequencer global parameters can be set up from the front panel from within the Sequence
Analyzer mode. To set these parameters proceed as follows:
Step 1. Press
Meas Setup, Result Type and set as required (currently, Measurement Metric is the
only supported type).
Step 2. Press
Meas Setup, Trigger Timeout and set as required (select On or Off, and set the time in
ms).
Step 3. Press
Meas Setup, Include Source in Sequence and set as required (Yes or No). This
setting is available only if
Source, List Sequencer, Sequencer is On.
Step 4. Press
Meas Setup, Abort on Limit Fail and toggle as required (On or Off).
Step 5. Press
Meas Setup, Abort on Error and toggle as required (On or Off).
Set up the input trigger which is either a signal burst at the input RF port or a signal at one of two input
trigger connectors on the back panel. Which trigger is used is specified in the acquisition setup.
Press
Meas Setup, Input Trigger Setup, Video, Trig Slope and toggle as required (Positive or
Negative).
or
Step 1. Press
Meas Setup, Input Trigger Setup, External 1, Trigger Level and set as required (level
in Volts).
Step 2. Press
Meas Setup, Input Trigger Setup, External 1, Trig Slope and set as required (Positive
or Negative).
or
Step 1. Press
Meas Setup, Input Trigger Setup, External 2, Trigger Level and set as required (level
in Volts).
Step 2. Press
Meas Setup, Input Trigger Setup, External 2, Trigger Slope and set as required
(Positive or Negative).
Set up the trigger holdoff, if required:
Press
Meas Setup, Input Trigger Setup, Trigger Holdoff, Trig Slope and toggle as required (Positive
or Negative).
Step 1. Press
Meas Setup, Input Trigger Setup, Trigger Holdoff, Trigger Holdoff and set as
required (time in seconds).
Step 2. Press
Meas Setup, Input Trigger Setup, Trigger Holdoff, Holdoff Type and set as required
(Below or Above).
Press
Meas Setup, Multiport Adaptr Input Mode and toggle as required (List or Fixed). If Fixed is
selected, the Multiport Adpater Input Port remains the same throughout the sequence, and is as defined
on the
Input/Output menu (see “Input/Output setup” on page 79). If List is selected, the Multiport
Adapter Input Port can be set independently for each acquisition, as defined by a sequence parameter
(see “Acquisition setup” on page 81).