Technical data

70 Chapter 5
Mode Parameters
Advanced Setup Parameters
Mode Parameters
TD-SCDMA Specifc Setup Parameters
Loopback BER Pattern, TD-SCDMA
This parameter allows you to specify (by file or by string) the reference bits used in TD-SCDMA
Loopback BER calculation.
:SERVice[:PRODuction]:LSEQuencer:ADVanced:TDSCdma:LBER:PATTern <filepath +
filename>
:SERVice[:PRODuction]:LSEQuencer:ADVanced:TDSCdma:LBER:PATTern:LOAD
<String>
:SERVice[:PRODuction]:LSEQuencer:ADVanced:TDSCdma:LBER:PATTern:LOAD?
Loopback BER CodeRMC, TD-SCDMA
This parameter allows you to specify the RMC (Reference Measurement Code) decoder type used in
TD-SCDMA Loopback BER calculation. Its type is bool. TRUE means it is single code, FALSE means
it is multi code.
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:RMC:SCODe TRUE|FALSE
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:RMC:SCODe?
Loopback BER CodenLevel, TD-SCDMA
This parameter allows you to specify the code level of the Reference Measurement Code used in
TD-SCDMA Loopback BER calculation.
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:CLEVel <int>
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:CLEVel?
Loopback BER Code Index0, TD-SCDMA
This parameter allows you to specify the code index of the Reference Measurement Code used in
TD-SCDMA Loopback BER calculation. If singleCodeRMC is true, it is only necessary to set this value;
if it is false, it is necessary to set this value and also the value of Loopback BER Code Index1.
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:CINDex0 <int>
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:CINDex0?
Loopback BER Code Index1, TD-SCDMA
This parameter allows you to specify the code index of the Reference Measurement Code used in
TD-SCDMA Loopback BER calculation. If singleCodeRMC is true, it is not necessary to set this value;
if it is false, it is necessary to set this value and also the value of Loopback BER Code Index0.
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:CINDex1 <int>
[:SENSe]:LSEQuencer:ADVanced:TDSCdma:LBER:CINDex1?