Technical data

68 Chapter 5
Mode Parameters
Advanced Setup Parameters
Mode Parameters
Advanced Setup Parameters
These are radio format specific parameters which apply to an entire sequence, and will apply to all
acquisitions and measurements related to that radio format.
GSM/EDGE Specifc Setup Parameters
Ignore Error In Average
If this parameter is ON, the application does not terminate the analysis step when an error is found. The
measurement is made, but all results with errors are excluded from the average. In this case, the actual
number of averages may be less than the requested number of averages, because some results were
excluded.
[:SENSe]:LSEQuencer:ASETup:GSM:IEAVerage ON|OFF|1|0
[:SENSe]:LSEQuencer:ASETup:GSM:IEAVerage?
PVT Time Offsets
The user may specify 12 offset time values for which power levels can be queried.
[:SENSe]:LSEQuencer:ASETup:GSM:PVTime:TIME[:OFFSet] <time>, <time>,
<time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>,
<time>
[:SENSe]:LSEQuencer:ASET:GSM:PVT:TIME[:OFFSet]?
Loopback BER Pattern, GSM
This parameter allows you to specify (by file or by string) the reference bits used in GSM Loopback
BER calculation.
:SERVice[:PRODuction]:LSEQuencer:ADVanced:GSM:LBER:PATTern <filepath +
filename>
:SERVice[:PRODuction]:LSEQuencer:ADVanced:GSM:LBER:PATTern:LOAD {1:4},
<String>
:SERVice[:PRODuction]:LSEQuencer:ADVanced:GSM:LBER:PATTern:LOAD? {1:4}
Loopback BER Pattern, EDGE
This parameter allows you to specify (by file or by string) the reference bits used in EDGE Loopback
BER calculation.
:SERVice[:PRODuction]:LSEQuencer:ADVanced:EDGE:LBER:PATTern <filepath +
filename>
:SERVice[:PRODuction]:LSEQuencer:ADVanced:EDGE:LBER:PATTern:LOAD {1:4},
<String>
:SERVice[:PRODuction]:LSEQuencer:ADVanced:EDGE:LBER:PATTern:LOAD? {1:4}