Technical data
Chapter 4 49
Coordinating the Sequencers
Include Source in Sequence is disabled
Coordinating the Sequencers
Once the analyzer list sequencer is initiated, the order of events is as follows:
• The controller begins the non-signalling test by instructing the DUT to transmit an RF signal.
• The analyzer list sequencer, triggered by the RF Burst received from the DUT, sends a trigger to the
source list sequencer.
• The source list sequencer begins executing its sequence and generating an RF output to the DUT.