Technical data

Chapter 3 41
Source List Sequencer Parameter Definitions
Source Parameters
Source List Sequencer Parameter
Definitions
10: Time / Count
This parameter specifies the duration of the step in ms (if the Step Duration parameter has been set to
“TIME”), or as the number of times a waveform file is played (if the Step Duration parameter has been
set to “COUNt”). If the Step Duration parameter has been set to “CONTinuous”, this parameter is
ignored.
The key path for selecting the Duration from the front panel menus is:
[Source], List Sequencer, List Sequencer Setup, Step Duration, Time/Count
11: Output Trigger
This parameter specifies the type of trigger output signal which will be sent at the start of the step. The
choices are:
ON: the trigger output is sent to the test set’s analyzer (for triggering of the analyzer list sequencer),
and also to the TRIGGER 2 OUT port at the rear panel of the test set.
OFF; no trigger output is generated by the source.
The key path for selecting Output Trigger from the front panel menus is:
[Source], List Sequencer, List Sequencer Setup, Output Trigger