Technical data

Chapter 16 249
1xEV-DO Programming Commands
Modulation Accuracy (Rho)
1xEV-DO Programming Commands
[:SENSe]:LSEQuencer:EVDO:RHO:MS:FERRor:TRANge
Modulation Accuracy results
Index Result Parameter
0 Overall Pass/Fail Result shows the mask test result (0.0 = pass, 1.0 = fail, -1.0:
Not tested)
1 EVM pass/fail result (1.0 = fail, 0.0 = pass)
2 Peak EVM pass/fail result (1.0 = fail, 0.0 = pass)
3 Rho pass/fail result (1.0 = fail, 0.0 = pass)
4 Peak Code Domain Error pass/fail result (1.0 = fail, 0.0 = pass)
5 Frequency Error pass/fail result (1.0 = fail, 0.0 = pass)
6 Pilot Offset pass/fail result (1.0 = fail, 0.0 = pass)
7 Max Inactive Channel Code Domain Power pass/fail result (1.0 = fail, 0.0 =
pass)
8 RRI Relative Power pass/fail result (1.0 = fail, 0.0 = pass)
9 ACK Channel Relative power pass/fail result (1.0 = fail, 0.0 = pass)
10 DRC Channel Relative power pass/fail result (1.0 = fail, 0.0 = pass)
11 Data Channel Relative power pass/fail result (1.0 = fail, 0.0 = pass)
12 DSC Channel Relative power pass/fail result (Subtype 2/3 only) (1.0 = fail, 0.0
= pass)
13 Auxiliary Pilot Channel Relative Power pass/fail result (Subtype 2/3 only) (1.0
= fail, 0.0 = pass)
14 Average RMS EVM – a floating point number (in percent) of EVM over the
entire measurement area.
15 Peak EVM– a floating point number (in percent) of peak EVM over the entire
measurement area.
16 Magnitude error– a floating point number (in percent) over the entire
measurement area.
17 Phase error– a floating point number (in degrees) of average phase error over
the entire measurement area.
18 I/Q Origin Offset – a floating point number (in dB) of the I and Q error
(magnitude squared) offset from the origin.
19 Frequency Error – a floating point number (in percent) of average magnitude
error over the entire measurement area.
20 Rho – a floating point number of Rho.
21 Peak Code Domain Error – a floating point number (in dB) of the Peak Code
Domain Error relative to the main power.