Technical data
172 Chapter 12
Basic Measurements Programming Commands
Basic Transmit Power
Basic Measurements Programming
Commands
Basic Transmit Power
The basic transmit power is a measure of the RMS or average power in the received RF signal for the
duration specified in the Analysis Interval.
To set up a basic transmit power measurement in the List Sequencer from the front panel do the
following:
Step 1. Set the test set to the Sequence Analyzer mode:
Press
Mode, Sequence Analyzer, Mode Preset.
Step 2. Set up the Basic Transmit Power parameters:
Press
Meas Setup, Basic Meas Setup, Basic Transmit Power Setup.
Press
Upper Limit to specify upper limit of power.
Press
Lower Limit to specify lower limit of power.
Press
Radio Standard [none], Digital IF BW and adjust as required.
Press
Filter Type and select as required (Gaussian, Flattop, Root Raised Cosine, or Short
Nyquist
).
Press
Filter BW and adjust as required (not applicable to flattop filtering).
Press
Filter Alpha and adjust as required (not applicable to flattop filtering).
Step 3. Set up an Acquisition in the test set. To do this see “Front Panel List Sequencer Setup” on
page 78)
Figure 12-1. Basic Transmit Power measurement