Technical data

Chapter 9 125
Sequence Setup Commands
Programming Analyzer Sequences (Parameter-Specific Commands)
Sequence Setup Commands
18: Acquisition Integration
For a description of this parameter, see “18: Acquisition Integration” on page 30. The parameter-specific
commands are described below.
Remote Command
To set the Acquisition Integration as an individual value:
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:INTegration
NORMal|PRIMary|LOWer|UPPer|RANGe
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:INTegration?
To set the Acquisition Integration as an array:
[:SENSe]:LSEQuencer:LIST:SETup:INTegration
<enum>,<enum>,<enum>,…
[:SENSe]:LSEQuencer:LIST:SETup:INTegration?
The array form of the command is used to make a different setting for each
acquisition in the sequence (the number of array values must match the
number of acquisitions; otherwise a "settings conflict" error will occur).
Example :LSEQ:ACQ2:SET:INT PRIM
:LSEQ:ACQ2:SET:INT?
:LSEQ:LIST:SET:INT PRIM,LOW,UPP,RANG
:LSEQ:LIST:SET:INT?