Technical data
11 0 Chapter 9
Sequence Setup Commands
Programming Analyzer Sequences (Combination Commands)
Sequence Setup Commands
Analysis Step Parameters (Combination)
The parameters for an analysis step within an acquisition can be set by means of a single SCPI
command, as described below. To set just one parameter (for a single analysis interval, or for all analysis
intervals in the acquisition at once), see “Analysis Step Parameters (Parameter-Specific Commands)” on
page 126.
Remote Command [:SENSe]:LSEQuencer:ACQuire[1]|2|3|4..512:ASTep[1]|2|3|
4…1000:SETup <time>, <time>, <integer>, <ampl>
Example :LSEQ:ACQ1:ASTEP1:SETup 20,20,1, –10
:LSEQ:ACQ1:ASTEP1:SETup?
Notes The parameters are:
<time> - specifies the Analysis Offset for the Analysis Step.
<time> - specifies the Analysis Interval for the Analysis Step.
<integer> - specifies the Measurement Bitmap for the Analysis Step (this is
the decimal equivalent of the bits selected).
<ampl> - specifies the Expected Power at DUT Output for the Analysis Step.