Technical data
Chapter 8 103
List Sequencer Measurement Results
Sequence Analyzer measurement results
List Sequencer Measurement Results
Results format
Assuming that the subopcode used with the he FETCh or READ command is 1 or none (indicating that
full results are requested), the results returned are as in the table below. In this table, each real value is a
“double” or 8 bytes (64 bits).
Step Result Comment
1 Total number of real values returned for the Sequence, including this one. Sequence as a
whole.
2 Reserved for Agilent use.
3 Summary Integrity value for entire Sequence. All integrity values OR’d together.
4 Total number of Acquisitions for entire Sequence.
5 Summary Integrity value for Acquisition 1. Acquisition 1
6 Number of Analysis Intervals in Acquisition 1
7
Summary Integrity value for Analysis Interval 1 Analysis Interval 1
of Acquisition 1
8
Analysis Interval 1 Measurement Bit Map - decimal representation
9
Integrity value for 1st measurement in Analysis Interval 1 Measurement 1 of
Analysis Interval 1
of Acquisition 1
10 Number of results in first measurement in Analysis Interval 1
11 Measurement results for first measurement in Analysis Interval 1
12
Integrity value for 2nd measurement in Analysis Interval 1 Measurement 2 of
Analysis Interval 1
of Acquisition 1
13 Number of results in 2nd measurement in Analysis Interval 1
14 Measurement results for 2nd measurement in Analysis Interval 1
15 ...repeat 12 to 14 for all other measurements in Analysis Interval 1 ...other
measurements of
Analysis Interval 1
of Acquisition 1
16
Summary Integrity value for Analysis Interval 2 Analysis Interval 2
of Acquisition 1
17
Analysis Interval 2 Measurement Bit Map - decimal representation
18
Integrity value for 1st measurement in Analysis Interval 2 Measurement 1 of
Analysis Interval 2
of Acquisition 1
19 Number of results in first measurement in Analysis Interval 2
20 Measurement results for first measurement in Analysis Interval 2