Technical data
72 Getting Started Guide
3 Test Set Operating System
It is possible a newer version of the instrument software is available on the web at
http://www.agilent.com/find/e6630a
4 After the E6630A restarts, it will boot up in
the default “Instrument” account, perform
alignments, and finally display the
measurement screen.
If the instrument does not reboot, remove
the USB storage device and cycle power.
The boot order for the instrument may
have been changed in the instrument
BIOS previously, and the instrument may
be instructed to boot from USB first, and
the internal drive second. If this is the
case, failure to remove the USB device
will result in an error, as the instrument
attempts to boot from it.
5 Check for any errors, and verify that the
new software version is installed.
In the right-click menu, Press System. In
the System menu, press
Show, System.
Look for the line “Instrument S/W
Revision”.
Instrument S/W Revision:
Additional recovery steps may be required to fully recover the system to a
more current working state. This could involve restoring your own backups of
the instrument configuration, including re-installing applications, data, and
performing system customizations.
Preparing the installation
Step Notes