Turning on the test set the first time
Initializing the test set using a USB mouse and external monitor
During the initial power-on process, the test set turns off and restarts
several times. This only happens during the initial installation process.
Steps Actions Notes
1 Connect a mouse and
keyboard
Connect the mouse and
keyboard to two of the USB
ports on the front panel
panel of the test set.
2 Connect a VGA monitor
Connect the monitor to the
monitor port on the front
panel of the test set.
Model E6630A has no display; an external monitor
is required.
See “Virtual Front Panel" on page 34.
3 Power on the test set a Position the test set so you
have easy access to the
power cord and plug it in.
b Press the power switch
(located near the top right
corner of the PXI rack) to
turn the test set on.
See “Test Set Location and Rack Mounting
Requirements and “Power requirements" on page 22
for more details.
The test set can require more than 5 minutes to
power-on. This is affected by Windows start-up
requirements.
The Agilent Technologies screen appears followed
by a screen that allows you to select the operating
system.
At this time, it is safe to turn off the test set before
starting the Setup Wizard.