Keysight Technologies

dataTec ▪  Ferdinand-Lassalle-Str. 52  ▪  72770 Reutlingen  ▪  Tel. 07121 / 51 50 50  ▪  Fax 07121 / 51 50 10  ▪  info@datatec.de  ▪  www.datatec.de
7
Product Specications
Denitions and conditions
Specication
Specications describe the performance parameters covered by the product warranty
and are valid from 20 to 35 °C unless otherwise noted.
Typical
Typical describes additional product performance information that is not covered by the
product warranty. It is performance beyond specications that 95 percent of the units
exhibit with a 95 percent condence level. This data, shown in italics, does not include
measurement uncertainty, and is valid only at room temperature (approximately 25 °C)
after alignment within the stated alignment time and temperature limits.
Nominal
Nominal values indicate expected performance, or describe product performance that is
useful in the application of the product, but are not covered by the product warranty.
The test set will meet its specication when:
The test set is within its calibration cycle
The test set has been stored at an ambient temperature within the allowed operat-
ing range for at least two hours before being turned on; if it has previously been
stored at a temperature range inside the allowed storage range, but outside the
allowed operating range
The test set has been turned on for at least 45 minutes
The RF, IF, and Source Alignments
1
have been run within the previous 7 days
An ALL Alignment
1
has been run:
Within the previous 8 hours
If the temperature has changed more than 5 °C from the previous
ALL” alignment
1. For more information on using alignments in a manufacturing environment, please see the EXM user documentation.
07 | Keysight | E6640A EXM Wireless Test Set - Data Sheet