7
Product Specications
Denitions and conditions
Specication
Specications describe the performance parameters covered by the product warranty
and are valid from 20 to 35 °C unless otherwise noted.
Typical
Typical describes additional product performance information that is not covered by the
product warranty. It is performance beyond specications that 95 percent of the units
exhibit with a 95 percent condence level. This data, shown in italics, does not include
measurement uncertainty, and is valid only at room temperature (approximately 25 °C)
after alignment within the stated alignment time and temperature limits.
Nominal
Nominal values indicate expected performance, or describe product performance that is
useful in the application of the product, but are not covered by the product warranty.
The test set will meet its specication when:
– The test set is within its calibration cycle
– The test set has been stored at an ambient temperature within the allowed operat-
ing range for at least two hours before being turned on; if it has previously been
stored at a temperature range inside the allowed storage range, but outside the
allowed operating range
– The test set has been turned on for at least 45 minutes
– The RF, IF, and Source Alignments
1
have been run within the previous 7 days
– An ALL Alignment
1
has been run:
– Within the previous 8 hours
– If the temperature has changed more than 5 °C from the previous
“ALL” alignment
1. For more information on using alignments in a manufacturing environment, please see the EXM user documentation.
07 | Keysight | E6640A EXM Wireless Test Set - Data Sheet