User`s guide

66 Chapter 3
Screen Reference
AUTOMATIC TEST
If the Screen Simp/Detai softkey is selected, the test items in the table below are
shown. Pass/Fail test results against the limits set on the Test Setup: Test Condition
screen in the CONFIGURATION mode are also shown.
TCH (Talk)
a
Specifies the traffic channel depending on the parameter selection
in the Variable field. The frequency corresponding to the channel
accompanies.
If TCH is selected in the Variable field, this specifies one traffic
channel to make Talk tests in addition to six traffic channels for RF
tests.
If DC Power is selected in the Variable field, this specifies one
traffic channel to make both Talk tests and RF tests.
Varia bl e
Selects either TCH or DC Power as follows:
o TCH: Enables to make six measurements on six traffic
channels specified in the TCH input fields of the Test
Item table, with or without one fixed dc voltage.
o DC Power: Enables to make six measurements on one
traffic channel with six dc voltages specified in the DCP
input fields of the Test Item table.
TCH or DCP
a
(up to 6)
According to the selection of Variable, either of the following
parameters need to be specified for a new test:
o TCH: Accepts up to six traffic channels with regard to
Multi Band. The allowable range is as follows:
GSM900: 1 to 124
E-GSM: 0 to 124, and 975 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810
o DCP: Accepts up to six dc voltages ranging from 3.0 to
11.0 V in 0.1 V steps.
a. The variable increment softkey is activated. Refer to "Variable Incre-
ment Softkey" on page 60.
b. When DC Power is set to Off, DC Current cannot be measured and
“----” is shown in the DC Current cells in the table of the AUTO-
MATIC TEST: Stand-by detailed screen.
Input Field Description