User`s guide
Chapter 2 31
Using Functions
Testing a Dual Band Mobile with AUTOMATIC TEST
2. Using Functions
Step 12. Press the Test Condition softkey to go to the next screen to set the detailed test
condition including high and low test limits as shown below.
Step 13. Set the fields of the test parameters as shown in the left side of the above screen
moving the circular cursor accordingly.
Step 14. Enter the appropriate limits in the cells of the test item table above.
Step 15. Repeat Step 13 to Step 14 to set the Test Condition screen by replacing GSM900
with DCS1800 in Radio Standard.
Step 16. After you have set all parameters as needed, press the Return, Return again, then
AUTOMATIC TEST softkeys to obtain the AUTOMATIC TEST: Stand-by
detailed screen shown below.
Step 17. Turn the mobile phone on. Wait for the mobile to camp on and display 001-01.
If you set DC Power or RF Output in the Test Sequence screen to Auto, turn the
mobile on after pressing the Start softkey.
Step 18. Press the Start softkey to begin your test.
The DC Current measurement at Camp On stage is measured at the LOC. Update
step and its pass/fail result is displayed in the DC Current cells of the table, when
DC Power is supplied from the Test Set to the mobile.
Step 14
Step 13