User`s guide

154 Chapter 3
Screen Reference
CONFIGURATION
Automatic Test Flow
For creating a test flow to execute AUTOMATIC TEST, refer to Table 3-1.
Select either MS or BS in the second step of the sequence, then the fifth, sixth and
seventh steps are automatically set accordingly.
Select Run to include a sequence in a test flow, or select --- (skip) to exclude that
sequence from a test flow.
Table 3-1 Setting an Automatic Test Flow
Place the cursor in front of the input field of a sequence and press the CURSOR
CONTROL knob one time to activate this input field. Select either MS or BS, for
example, by rotating the CURSOR CONTROL knob, and set it by pressing the
knob.
Step Sequence Action Test Action
1
LOC. Update:
Run|--- <Select either Run or --->
2
MS|BS Call:
<Select either MS or BS> Run
3
Talk:
Run|--- <Select either Run or --->
4
RF Test:
Run
5
MS|BS Release:
<Determined by Step 2> Run
6
BS|MS Call:
<Determined by Step 2> Run|--- <Select either Run or --->
7
BS|MS Release:
<Determined by Step 2> Run|--- <Determined by Step 6>