Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 95
Programming Command Reference
TESTs Subsystem for MANUAL TEST
3. Programming Command
Reference
This command specifies a voltage value of DC Power when the DC Power supply
mode in the Stand-by screen is set to ON. The allowable range is from 3.0 to 11.0
(V) in 0.1 steps. At preset, this is set to 3.0 (V).
Midamble
TESTs:MANual:TPOWer:MIDamble
NONE|TSC0|TSC1|TSC2|TSC3|TSC4|TSC5|TSC6|TSC7
TESTs:MANual:TPOWer:MIDamble?
This command specifies a value of Training Sequence Code (TSC), TSC0 to
TSC7, or NONE for the amplitude trigger. At preset, this is set to TSC5.
NOTE For Test Sets without Option 002 or if the mode is set to SYNC, this
command is invalid.
For Test Sets with Option 040, if the mode is set to GPRS, this
command is invalid.
Peak TX Power Measurement
TESTs:MANual:TXAMplitude:RESults:TPOWer?
This query command returns a set of values, <string 1>, <string 2> and
<string 3>, for Peak TX Power tests.
• <string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.
• <string 2> (1 digit) is the binary data, 0 or 1, for the pass (0) or fail (1) result.
• <string 3> (5 digits) is the measured power level in dBm ranging from −99.9 to
+99.9. If its measurement is overranged, underranged or not made, the value
−99.9 is returned.
Power Control Level
TESTs:MANual:TPOWer:PLEvel <int>
TESTs:MANual:TPOWer:PLEvel?
This command specifies a value of Power Control Level for each radio standard set
by the Radio Format command. The allowable range and preset values are as
follows:
RFORmat Allowable Range Preset Value
GSM900 2 (+39 dBm) to 19 (+5 dBm) 5 (+33 dBm)