Specifications

Table Of Contents
90 Chapter 3
Programming Command Reference
TESTs Subsystem for MANUAL TEST
Signaling State (GPRS mode)
TESTs:MANual:MEASure:SIGNaling:STATe?
This query command returns a value, <string>, corresponding to the test flow
steps executed for a test cycle for the GPRS mode as follows:
bit 0 (1): Idle
bit 1 (2): Attached
bit 2 (4): Connect
bit 3 (8): Time Out Error
bit 4 (16): Measurement End
The choices for each bit are 1 for executed and 0 for not.
NOTE If the mode is set to SYNC on the Test Set with Option 040, this
command is invalid.
Signaling State (SYNC mode)
TESTs:MANual:MEASure:SIGNaling:STATe?
This query command returns a value, <string>, corresponding to the test flow
steps executed for a test cycle as follows:
bit 0 (1): MS Call/BS Call
bit 1 (2): Talk/RF Test
bit 2 (4): MS Release/BS Release
bit 3 (8): Time Out Error
bit 4 (16): Measurement End
The choices for each bit are 1 for executed and 0 for not.
NOTE For Test Sets with Option 002, if the mode is set to ASYNC, this
command is invalid.
If the mode is set to GPRS on the Test Set with Option 040, this
command is invalid.
Slot Configuration
TESTs:MANual:MEASure:SCONfigure 1×1|2×1
This command specifies the slot combination for downlink and uplink,
respectively, in the Measuring screen for the GPRS mode, in conjunction with the
Slot Measured setting. The combinations of the slot configuration and the slot
order measured are as follows: